Search...

Portable VOC Detector

IVOC-3000
IVOC-3000

Portable VOC Detector

IVOC-3000 portable VOCs detector is a VOCs total amount (total length) detector through explosion-proof certification, the use of FID detector (hydrogen flame ionization detector) can well meet the needs of customers for a variety of on-site rapid and accurate detection of VOCs total amount (total vertical >). The product has the characteristics of small size, light weight, good detection performance, wide range and simple operation.

Inquire Now

FEATURES

·Hydrogen flame ionization detector (FID), it is a typical destructive, mass detector, is the flame generated by hydrogen and air combustion as energy, when organic compounds enter the flame of hydrogen and oxygen combustion, chemical ionization is produced at high temperature, ionization produces ions several orders of magnitude higher than the base flow, under the directional action of high voltage electric field, the formation of ion flow, The weak ion stream is amplified with high resistance and becomes an electrical signal proportional to the amount of organic compounds entering the flame, so organic matter can be quantitatively analyzed according to the size of the signal.

·Application Fields:

Petrochemical,chemical,pharmaceutical,resin,fertilizerand other industries; LDAR (Leak Detection and Repair);Soil VOCs detection; Factory environment vocs detection;Fugitive emission field emergency detection; Rapiddetection of emissions from stationary pollution sources;Effect evaluation of vocs treatment facilities; vocssource investigation; Fugitive vocs detection ofwastewater pools; Oil gas recovery inspection at gasstations or bulk petroleum terminals.

·Manual operator

The hand operator is composed of explosion-proof smart phone and portable VOCs analysis software. The hand operator communicates with the instrument through Bluetooth, and the APP realizes Bluetooth connection, ignition, calibration, reading, result recording and query.


SPECIFICATIONS

ModelIVOC-3000
Detection limit 0.5ppm
Measuring range 0-50000ppm
Accuracy <±5% (methane standard gas :500ppm)
Response time T90 in 3.5 seconds (1 meter long sample tube)
Hydrogen bottle Capacity 100ml, maximum working pressure 20MPa Hydrogen use time: more than 12 hours
Battery working time 10h
Weight 3.85Kg
Shell size 25.0*19.3*6.2 (cm)
Explosion-proof mark Ex d ia mb llCT4 Gb
Ambient temperature  -10℃~+45℃
Repeatability <2% (methane standard gas :500ppm)


BROCHURE

Soil VOC detector

File Type:.pdf File Size:1.39 MB

INQUIRE NOW

0086-532-87890634